Are Standard Human Coagulation Tests Suitable in Pigs and Calves During Extracorporeal Circulation?
- 1 July 2001
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 25 (7) , 579-584
- https://doi.org/10.1046/j.1525-1594.2001.025007579.x
Abstract
The thrombogenicity of membrane oxygenators as well as clotting parameters profiles, using standard human clotting tests, was analyzed in calves and pigs during 6 h perfusion. Three calves and 3 pigs were connected to extracorporeal circulation with standard heparinization. Blood samples were taken for coagulation variables throughout perfusion, and oxygenators were examined for clot deposits at the end of the experiment. Two out of 3 oxygenators of the calf group presented clot deposits while none in the pig group did. Baseline coagulation variables of pigs showed values similar to those of humans while neither extrinsic nor intrinsic pathways could be activated in calves with standard human coagulation tests. The calf model, in conclusion, was confirmed to be a difficult model for the testing of extracorporeal circulation device resistance to thrombus formation, which is, however, not reflected by standard human coagulation tests. The pig model is a better model in which both coagulation pathways could be activated with standard human coagulation tests.Keywords
This publication has 21 references indexed in Scilit:
- A Siliconized Hollow Fiber Membrane OxygenatorAsaio Journal, 2000
- The platelet in cardiopulmonary bypassThe Annals of Thoracic Surgery, 1998
- Experimental Evaluation of the Medtronic Maxima Forte Hollow Fiber Membrane OxygenatorThe Journal of ExtraCorporeal Technology, 1998
- Thromboembolic Phenomena and the use of the Pig as an Appropriate Animal Model for Research on Cardiovascular DevicesThe International Journal of Artificial Organs, 1997
- Platelet Glycoprotein IIb/IIIa Receptors in Cardiovascular MedicineNew England Journal of Medicine, 1995
- Low prime, heparin coated centrifugal pumpsPerfusion, 1995
- Experimental Evaluation of Heparin-Coated Cardiopulmonary Bypass Equipment with Low Systemic Heparinization and High-Dose AprotininThe Thoracic and Cardiovascular Surgeon, 1991
- Heparin-coating of extracorporea circuits reduces thrombin formation in patients undergoing cardiopulmonary bypassPerfusion, 1991
- Long Term Cardiopulmonary Bypass without Systemic HeparinizationThe International Journal of Artificial Organs, 1990