Wear‐accelerated corrosion of Ti‐6Al‐4V and nitrogen‐ion‐implanted Ti‐6Al‐4V: Mechanisms and influence of fixed‐stress magnitude
- 1 March 1987
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 21 (3) , 367-377
- https://doi.org/10.1002/jbm.820210309
Abstract
Wear-accelerated corrosion rates at constant anodic potentials were evaluated for unimplanted and nitrogen-ion-implanted surgical Ti-6Al-4V while wearing against ultrahigh-molecular-weight polyethylene at stress levels up to 6.90 MPa (1000 psi). The ion implantation processing was found to reduce the wear corrosion rates in both saline and serum solutions at all applied stress levels. During wear testing, all of the ion-implanted surfaces remained visually unchanged from the polished condition. However, many of the unimplanted surfaces developed damage zones characterized by wear tracks and black wear debris. A surface-damage mechanism is proposed and discussed which involves disruption of the Ti-6Al-4V protective oxide film, subsequent entrapment of oxide particles in the polyethylene, then self-perpetuating damage due to the abrasive action of the embedded particles.This publication has 5 references indexed in Scilit:
- Ion implantation of surgical Ti6Al4V alloyMaterials Science and Engineering, 1985
- A study of the improved wear performance of nitrogen-implanted Ti-6Al-4VWear, 1983
- Technical Note:A New Apparatus for Synergistic Studies of Corrosive WearCorrosion, 1983
- Practical Applications of Ion ImplantationJOM, 1982
- Wear and its measurement for joint prosthesis materialsWear, 1978