Pitfalls Using Doppler Ultrasound to Transduce Blood Flow
- 1 July 1973
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-20 (4) , 306-309
- https://doi.org/10.1109/tbme.1973.324201
Abstract
General testing procedures which will ascertain whether a Doppler ultrasonic blood flowmeter is correctly designed and operating properly are presented. A commercial Doppler ultrasonic blood flowmeter using the procedure is evaluated. Two pitfalls were encountered which prevented its proper operation. High-frequency response was inadequate, which introduced a nonlinearity into the input-output calibration curve. Low-frequency gain was too high, resulting in wall-motion artifacts. These problem areas are described in detail, and a remedy for each is suggested.Keywords