Spiral scanning with anisotropic field of view
- 1 March 1998
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 39 (3) , 448-456
- https://doi.org/10.1002/mrm.1910390315
Abstract
Spiral scanning has been used to achieve much shorter scan times than conventional techniques for a wide range of applications. The major drawback with spiral scans is blurring from off‐resonant spins, which is proportional to the readout time. Blurring limits maximal spatial resolution and minimal scan time potentially achievable with spiral scanning. Anisotropic field of view is used in conventional scanning to improve image quality by matching /c‐space trajectory to object characteristics. Anisotropic field of view improves spatial resolution in spiral scanning without increasing scan time or blurring. The resolution improvement results from increased maximal k‐space radius allowed by the lower field of view. A field of view reduction by a factor of 2 in one direction provides up to 60% resolution improvement in that direction. Reduced SNR also results from non‐uniform /c‐space sampling.Keywords
This publication has 17 references indexed in Scilit:
- Decomposition of inflow and blood oxygen level‐dependent (BOLD) effects with dual‐echo spiral gradient‐recalled echo (GRE) fMRIMagnetic Resonance in Medicine, 1996
- Spiral imaging on a small‐bore system at 4.7tMagnetic Resonance in Medicine, 1995
- Optimized gradient waveforms for spiral scanningMagnetic Resonance in Medicine, 1995
- Fast Three Dimensional Magnetic Resonance ImagingMagnetic Resonance in Medicine, 1995
- Magnetic resonance velocity imaging using a fast spiral phase contrast sequenceMagnetic Resonance in Medicine, 1994
- Real time blood flow imaging by spiral scan phase velocity mappingMagnetic Resonance in Medicine, 1994
- Fast Spiral Coronary Artery ImagingMagnetic Resonance in Medicine, 1992
- Deblurring for non‐2D fourier transform magnetic resonance imagingMagnetic Resonance in Medicine, 1992
- A homogeneity correction method for magnetic resonance imaging with time-varying gradientsIEEE Transactions on Medical Imaging, 1991
- Selection of a convolution function for Fourier inversion using gridding (computerised tomography application)IEEE Transactions on Medical Imaging, 1991