Calibrating an area-detector diffractometer: integral response
- 8 March 1990
- journal article
- Published by The Royal Society in Proceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences
- Vol. 428 (1874) , 181-214
- https://doi.org/10.1098/rspa.1990.0030
Abstract
The quality of diffraction data measured with electronic area-detectors is improved by correcting for non-uniformities in the response of the detector. Many detectors are actually much more uniform than they appear because much of the perceived non-uniformity is an artefact of the distortions in their imaging geometry and of the methods of illumination during calibration. Indeed, every known correction reduces the perceived non-uniformity. Our inability to illuminate the detector uniformly with radiation of the same wavelength as is used during data-collection is a particular worry because of differential absorption. The tails of the point-spread function also perturb the apparent response, particularly near to the edges of the imaging area. These problems are difficult to compensate, so there is no completely satisfactory method of determining the true response of real detectors. However, this does not prevent us from making calibrations of usable accuracy. Although this paper applies to all types of area-detector, the discussion is centred mainly on the ENRAF-NONIUS fast system, which is a commercially available television diffractometer, calibrated using software written by the present author. Calibrating the response of imaging detectors is a general problem, and many of the techniques expounded here are of wide applicability.This publication has 8 references indexed in Scilit:
- Calibrating an area-detector diffractometer: imaging geometryProceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences, 1989
- Miniature ionization chambersJournal of Applied Crystallography, 1988
- THE DEVELOPMENT OF A FULL PROFILE ANALYSIS OF SINGLE-CRYSTAL X-RAY DIFFRACTION DATALe Journal de Physique Colloques, 1986
- X-ray television area detectorsNuclear Instruments and Methods in Physics Research, 1982
- A New Method of Spectroscopy for Faint X-RadiationsJournal of the Optical Society of America, 1928
- XCIII. On the theory of X-ray absorption and of the continuous X-ray spectrumJournal of Computers in Education, 1923
- A New Method of X-Ray Crystal AnalysisPhysical Review B, 1917
- On X-Ray Wave-LengthsPhysical Review B, 1915