Rapid Small-Animal Dual-Energy X-Ray Absorptiometry Using Digital Radiography
- 1 December 2000
- journal article
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 15 (12) , 2451-2457
- https://doi.org/10.1359/jbmr.2000.15.12.2451
Abstract
Although dual-energy X-ray absorptiometry (DEXA) is an established technique for clinical assessment of areal bone mineral density (BMD), the spatial resolution, signal-to-noise ratio, scan time, and availability of clinical DEXA systems may be limiting factors for small-animal investigations using a large number of specimens. To avoid these limitations, we have implemented a clinical digital radiography system to perform rapid area DEXA analysis on in vitro rat bone specimens. A crossed step-wedge (comprised of epoxy-based materials that mimic the radiographic properties of tissue and bone) was used to calibrate the system. Digital radiographs of bone specimens (pelvis, spine, femur, and tibia from sham-ovariectomized [SHAM] and ovariectomized [OVX] rats) were obtained at 40 kilovolt peak (kVp) and 125 kVp, and the resulting areal BMD values were compared with those obtained with a clinical fan-beam DEXA system (Hologics QDR 4500). Our investigation indicates that the cross-wedge calibrated (CWC) DEXA technique provides high-precision measurements of bone mineral content (BMC; CV = 0.6%) and BMD (CV = 0.8%) within a short acquisition time (<30 s). Areal BMD measurements reported by the CWC-DEXA system are within 8.5% of those reported by a clinical fan-beam scanner, and BMC values are within 5% of the known value of test specimens. In an in vivo application, the CWC-DEXA system is capable of reporting significant differences between study groups (SHAM and OVX) that are not reported by a clinical fan-beam DEXA system, because of the reduced variance and improved object segmentation provided by the CWC-DEXA system.Keywords
This publication has 16 references indexed in Scilit:
- Bone Mineral Measurement of Phalanges: Comparison of Radiographic Absorptiometry and Area Dual X-ray AbsorptiometryRadiology, 2000
- Histo-anatomy of the proximal femur in rats: Impact of ovariectomy on bone mass, structure, and stiffnessThe Anatomical Record, 1996
- Effects of hysterectomy on bone in intact rats, ovariectomized rats, and ovariectomized rats treated with estrogenJournal of Bone and Mineral Research, 1996
- The Use of Dual-Energy X-ray Absorptiometry in AnimalsInvestigative Radiology, 1996
- The effect of ovariectomy on bone metabolism in ratsBone, 1995
- Dual-energy X-ray absorptiometry for measuring total bone mineral content in the rat: Study of accuracy and precisionBone and Mineral, 1994
- Precision of measurement by dual-energy X-ray absorptiometry of bone mineral density and content in rat hindlimb in vitroJournal of Bone and Mineral Research, 1994
- Dual-energy x-ray absorptiometry of the rat: Accuracy, precision, and measurement of bone lossJournal of Bone and Mineral Research, 1993
- Sequential and precise in vivo measurement of bone mineral density in rats using dual-energy x-ray absorptiometryJournal of Bone and Mineral Research, 1992
- Bone Density and CompositionJournal of Bone and Joint Surgery, 1966