On the field-size dependence of relative output from a linear accelerator
- 1 November 1992
- journal article
- Published by Wiley in Medical Physics
- Vol. 19 (6) , 1441-1444
- https://doi.org/10.1118/1.596799
Abstract
The radiation output in air on the central axis of a linac photon beam has been modeled as the sum of two components. These are a point source representing radiation direct from the target and a distributed source representing scatter in the flattening filter and primary collimator. By fitting only two parameters, the ratio of the two components for a 20 x 20 field and a width parameter for the distributed source this semi-empirical model describes the relative outputs of square, symmetric rectangular, and asymmetric rectangular fields with an average error of 0.25% for the field sizes studied.Keywords
This publication has 0 references indexed in Scilit: