The need for quality control in measurement of mixed venous oxygen saturation
- 1 October 1994
- journal article
- clinical trial
- Published by Wiley in Anaesthesia
- Vol. 49 (10) , 886-888
- https://doi.org/10.1111/j.1365-2044.1994.tb04267.x
Abstract
We investigated the variability of arterial, mixed venous and peripheral venous oxygen saturation readings in three co-oximeters in regular use on two Intensive Care Units. Over a 2-week period, 96 readings were obtained from each of the following machines; two AVL 912 co-oxylite co-oximeters (Machines A and B, Biomedical Instruments, Graz, Austria) and one IL-282 (Machine C, Instrumentation Laboratories, Lexington, MA, USA). The coefficient of variation in the arterial oxygen saturations was extremely small on all three machines (< 0.43%). The coefficient of variation in the mixed venous samples, however, ranged from 1.96 to 4.61% on machine A, and from 2.73 to 4.71% on machine B, but only from 0.17 to 1.47% on machine C. The variation in mixed venous saturations obtained from a single blood sample repeatedly analysed on machines A and B was large enough to influence clinical management.Keywords
This publication has 6 references indexed in Scilit:
- Arterial oxygen status determined with routine pH/blood gas equipment and multi-wavelength hemoximetry: Reference values, precision, and accuracyScandinavian Journal of Clinical and Laboratory Investigation, 1990
- Reliability of the determination of whole-blood oxygen affinity by means of blood-gas analyzers and multi-wavelength oximeters.Clinical Chemistry, 1989
- Bias and Precision of Pulse Oximeters and Arterial OximetersChest, 1988
- Continuous Venous Oximetry in Surgical PatientsAnnals of Surgery, 1986
- Therapy of Critically Ill Postoperative Patients Based on Outcome Prediction and Prospective Clinical TrialsSurgical Clinics of North America, 1985