Rotating Disk Potentiometry for Inner Solution Optimization of Low-Detection-Limit Ion-Selective Electrodes
- 31 October 2003
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 75 (24) , 6922-6931
- https://doi.org/10.1021/ac0346961
Abstract
The extent of optimization of the lower detection limit of ion-selective electrodes (ISEs) can be assessed with an elegant new method. At the detection limit (i.e., in the absence of primary ions in the sample), one can observe a reproducible change in the membrane potential upon alteration of the aqueous diffusion layer thickness. This stir effect is predicted to depend on the composition of the inner solution, which is known to influence the lower detection limit of the potentiometric sensor dramatically. For an optimized electrode, the stir effect is calculated to be exactly one-half the value of the case when substantial coextraction occurs at the inner membrane side. In contrast, there is no stir effect when substantial ion exchange occurs at the inner membrane side. Consequently, this experimental method can be used to determine how well the inner filling solution has been optimized. A rotating disk electrode was used in this study because it provides adequate control of the aqueous diffusion layer thickness. Various ion-selective membranes with a variety of inner solutions that gave different calculated concentrations of the complex at the inner membrane side were studied to evaluate this principle. They contained the well-examined silver ionophore O,O‘ ‘-bis[2-(methylthio)ethyl]-tert-butylcalix[4]arene, the potassium ionophore valinomycin, or the iodide carrier [9]mercuracarborand-3. Stir effects were determined in different background solutions and compared to theoretical expectations. Correlations were good, and the results encourage the use of such stir-effect measurements to optimize ISE compositions for real-world applications. The technique was also found to be useful in estimating the level of primary ion impurities in the sample. For an iodide-selective electrode measured in phosphoric acid, for example, apparent iodide impurity levels were calculated as 5 × 10-10 M.Keywords
This publication has 21 references indexed in Scilit:
- Rotating Ion‐Selective Membrane Electrodes for Trace‐Level MeasurementsElectroanalysis, 2003
- Improving the Detection Limit of Anion-Selective Electrodes: An Iodide-Selective Membrane with a Nanomolar Detection LimitAnalytical Chemistry, 2003
- Potentiometric Cd2+-selective electrode with a detection limit in the low ppt rangeAnalytica Chimica Acta, 2001
- Improved Detection Limits and Unbiased Selectivity Coefficients Obtained by Using Ion-Exchange Resins in the Inner Reference Solution of Ion-Selective Polymeric Membrane ElectrodesAnalytical Chemistry, 2000
- Effects of Ion Transport on the Potential Response of Ionophore-Based Membrane Electrodes: A Theoretical ApproachThe Journal of Physical Chemistry B, 1999
- Lowering the Detection Limit of Solvent Polymeric Ion-Selective Electrodes. 1. Modeling the Influence of Steady-State Ion FluxesAnalytical Chemistry, 1999
- Lowering the Detection Limit of Solvent Polymeric Ion-Selective Membrane Electrodes. 2. Influence of Composition of Sample and Internal Electrolyte SolutionAnalytical Chemistry, 1999
- Effect of Transmembrane Electrolyte Diffusion on the Detection Limit of Carrier-Based Potentiometric Ion SensorsAnalytical Chemistry, 1998
- Large Improvement of the Lower Detection Limit of Ion-Selective Polymer Membrane ElectrodesJournal of the American Chemical Society, 1997
- A Hexamethyl Derivative of [9]Mercuracarborand-3: Synthesis, Characterization, and Host−Guest ChemistryJournal of the American Chemical Society, 1996