MRI monitoring of heating produced by ultrasound absorption in the skull: In vivo study in pigs
Open Access
- 26 April 2004
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 51 (5) , 1061-1065
- https://doi.org/10.1002/mrm.20043
Abstract
The purpose of this study was to test the utility of MR thermometry for monitoring the temperature rise on the brain surface and in the scalp induced by skull heating during ultrasound exposures. Eleven locations in three pigs were targeted with unfocused ultrasound exposures (frequency = 690 kHz; acoustic power = 8.2–16.5 W; duration = 20 s). MR thermometry (a chemical shift technique) showed an average temperature rise in vivo of 2.8°C ± 0.6°C and 4.4°C ± 1.4°C on the brain surface and scalp, respectively, at an acoustic power level of 10 W. The temperature rise on the scalp agreed with that measured with a thermocouple probe inserted adjacent to the skull (average temperature rise = 4.6°C ± 1.0°C). Characterization of the transducer showed that the average acoustic intensity was 1.3 W/cm2 at an acoustic power of 10 W. The ability to monitor the temperature rise next to the skull with MRI-based thermometry, as shown here, will allow for safety monitoring during clinical trials of transcranial focused ultrasound. Magn Reson Med 51:1061–1065, 2004.Keywords
This publication has 18 references indexed in Scilit:
- Experimental demonstration of noninvasive transskull adaptive focusing based on prior computed tomography scansThe Journal of the Acoustical Society of America, 2003
- A non-invasive method for focusing ultrasound through the human skullPhysics in Medicine & Biology, 2002
- Micro-receiver guided transcranial beam steeringIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2002
- A hemisphere array for non-invasive ultrasound brain therapy and surgeryPhysics in Medicine & Biology, 2000
- Trans-skull ultrasound therapy: the feasibility of using image-derived skull thickness information to correct the phase distortionIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1999
- The potential of transskull ultrasound therapy and surgery using the maximum available skull surface areaThe Journal of the Acoustical Society of America, 1999
- Ultrasonic beam focusing through tissue inhomogeneities with a time reversal mirror: application to transskull therapyIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1996
- Transkull transmission of an intense focused ultrasonic beamUltrasound in Medicine & Biology, 1977
- Intense Ultrasound in Investigations of the Central Nervous SystemPublished by Elsevier ,1959
- A NEW METHOD FOR THE GENERATION AND USE OF FOCUSED ULTRASOUND IN EXPERIMENTAL BIOLOGYThe Journal of general physiology, 1942