Design and characterization of a digital image acquisition system for whole-specimen breast histopathology
- 22 September 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (20) , 5089-5103
- https://doi.org/10.1088/0031-9155/51/20/002
Abstract
We have developed a digital histopathology imaging system capable of producing a three-dimensional (3D) representation of histopathology from an entire lumpectomy specimen. The system has the potential to improve the accuracy of surgical margin assessment in the treatment of breast cancer by providing finer sampling and 3D visualization. A scanning light microscope was modified to allow digital photomicrography of a stack of large (up to 120 x 170 mm(2)) histology slides cut serially through the entire specimen. The images are registered and displayed in 2D and 3D. The design of the system, which reduces or eliminates the appearance of 'tiling' and 'seam' artefacts inherent in the scanning method, is described and its resolution, contrast/noise and coverage properties are characterized through measurements of the modulation transfer function (MTF), depth of field (DOF) and signal difference to noise ratio (SDNR). The imaging task requires a lateral resolution of 5 mu m, an SDNR of 5 between relevant features, 'tiling artefact' at a level below the detectability threshold of the eye, and 'seam artefact' of less than 5-10 mu m. The tests demonstrate that the system is largely adequate for the imaging task, although further optimizations are required to reduce the degradation of coverage incurred by seam artefact.Keywords
This publication has 22 references indexed in Scilit:
- Correlations between access to mammography and breast cancer stage at diagnosisCancer, 2005
- The prognostic influence of multifocality in breast cancer patientsThe Breast, 2004
- Escaping from Flatland: clinical and biological aspects of human mammary duct anatomy in three dimensionsThe Journal of Pathology, 2004
- Automatic segmentation of histological structures in mammary gland tissue sectionsJournal of Biomedical Optics, 2004
- Risk factors for local recurrence after breast-conserving surgeryBritish Journal of Surgery, 2003
- The Influence of Margin Width on Local Control of Ductal Carcinoma in Situ of the BreastNew England Journal of Medicine, 1999
- A method for measuring the presampled MTF of digital radiographic systems using an edge test deviceMedical Physics, 1998
- Three dimensional anatomy of complete duct systems in human breast: pathological and developmental implications.Journal of Clinical Pathology, 1996
- Objective image alignment for three-dimensional reconstruction of digital autoradiogramsJournal of Neuroscience Methods, 1988