Effect of the thermodynamics of an infant plethysmograph on the measurement of thoracic gas volume

Abstract
Adult plethysmographs have frequency responses that are essentially flat over the range of frequencies encountered in the measurement of thoracic gas volume (TGV). An infant plethysmograph is necessarily much smaller than an adult model. This means that there is a smaller mean distance over which heat diffusion must occur between the air in the plethysmograph and its walls. This in turn leads to a much reduced thermal time constant. We examined the effects of thermal time constant of a 60 L infant plethysmograph on measurements of TGV in infants. The thermal time constant was measured by rapidly injecting 20 mL of air into the plethysmograph, and found to be 0.16 ± 0.09 s when the box was empty. We calculated from this time constant that measurements of TGV should be quite dependent on the frequency at which the associated panting maneuvers are performed. TGV was measured in 5 infants less than 6 months old in the recovery phase following acute viral bronchiolitis. When we performed a digital correction of the measurements, to compensate for the thermal time constant of the plethysmograph, the TGV values decreased by a mean of 12%. Agitating the air in the plethysmograph with a fan decreased the thermal time constant of the box and reduced measured TGV by a mean of 8.4%. These results indicate that thermodynamics of infant plethysmographs can be an important source of error in TGV measurements. Pediatr Pulmonol 1990; 8:203‐208.