Gradient Elution in Capillary Electrochromatography
Open Access
- 1 September 1996
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 68 (17) , 2726-2730
- https://doi.org/10.1021/ac960020c
Abstract
In analogy to pressure-driven gradient techniques in high-performance liquid chromatography, a system has been developed for delivering electroosmotically driven solvent gradients for capillary electrochromatography (CEC). Dynamic gradients with submicroliter per minute flow rates are generated by merging two electroosmotic flows that are regulated by computer-controlled voltages. These flows are delivered by two fused-silica capillary arms attached to a T-connector, where they mix and then flow into a capillary column that has been electrokinetically packed with 3-μm reversed-phase particles. The inlet of one capillary arm is placed in a solution reservoir containing one mobile phase, and the inlet of the other is placed in a second reservoir containing a second mobile phase. Two independent computer-controlled, programmable, high-voltage power supplies (0−50 kV)one providing an increasing ramp and the other providing a decreasing rampare used to apply variable high-voltage potentials to the mobile phase reservoirs to regulate the electroosmotic flow in each arm. The ratio of the electroosmotic flow rates between the two arms is changed with time according to the computer-controlled voltages to deliver the required gradient profile to the separation column. Experiments were performed to confirm the composition of the mobile phase during a gradient run and to determine the change of the composition in response to the programmed voltage profile. To demonstrate the performance of electroosmotically driven gradient elution in CEC, a mixture of 16 polycyclic aromatic hydrocarbons was separated in less than 90 min. This gradient technique is expected to be well-suited for generating not only solvent gradients in CEC but also other types of gradients, such as pH and ionic strength gradients, in capillary electrokinetic separations and analyses.Keywords
This publication has 24 references indexed in Scilit:
- Capillary electrochromatography: operating characteristics and enantiomeric separationsJournal of Chromatography A, 1996
- Capillary Electrochromatography: Analysis of Polycyclic Aromatic HydrocarbonsAnalytical Chemistry, 1995
- Packed-capillary electrochromatographic separation of the enantiomers of neutral and anionic compounds using β-cyclodextrin as a chiral selector: Effect of operating parameters and comparison with free-solution capillary electrophoresisJournal of Chromatography A, 1994
- Ultrasensitive fluorescence detection of polycyclic aromatic hydrocarbons in capillary electrophoresisAnalytical Chemistry, 1993
- pH gradient capillary zone electrophoresis using a solvent program delivery systemAnalytical Chemistry, 1992
- Low-Cost Liquid Chromatography (LC-LC). IV. “Pulsed Open Tube Gradient Generation”, A New Approach for Generating Nanoliter Volume Linear and Tailored Gradients for Capillary Electrophoresis and Micro-LCJournal of Liquid Chromatography, 1990
- Step change of counterion – a new option in capillary zone electrophoresisElectrophoresis, 1989
- Gradient elution for micellar electrokinetic capillary chromatographyAnalytical Chemistry, 1988
- Electrochromatography using high applied voltageAnalytical Chemistry, 1987
- High-resolution separations based on electrophoresis and electroosmosisJournal of Chromatography A, 1981