Interferometers vs Monochromators: Separating the Optical and Digital Advantages
- 1 November 1977
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 31 (6) , 485-495
- https://doi.org/10.1366/000370277774464048
Abstract
Fourier transform (FT) and grating infrared spectrophotometers are compared from a theoretical viewpoint in terms of the multiplex advantage of FT spectrometers, the relative optical throughputs, the comparative performance of the detectors used with each type of system, and other more minor parameters. The signal/noise ratios of spectra measured using commercial FT and grating spectrophotometers are compared. The relative performance varies with wavenumber in the manner forecast theoretically, although the experimental performance of the FT spectrometer is slightly worse than we predicted. Optimum applications for each type of instrument are discussed in terms of these results.Keywords
This publication has 5 references indexed in Scilit:
- Infrared spectrophotometric observations of the adsorption of fibrinogen from solution at optically transparent carbon film electrode surfacesAnalytical Chemistry, 1976
- Fourier and Hadamard transform spectrometers: a limited comparisonApplied Optics, 1976
- Noise and Finite Register Effects in Infrared Fourier Transform SpectroscopyApplied Spectroscopy, 1976
- Errors in absorbance measurements in infrared Fourier transform spectrometry because of limited instrument resolutionAnalytical Chemistry, 1975
- Photometric Precision in Infrared Spectra Measured by Fourier Transform SpectroscopyApplied Spectroscopy, 1975