Calibration of Coulter Counters for Particles ∼1 μ in Diameter
- 1 November 1966
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 37 (11) , 1515-1520
- https://doi.org/10.1063/1.1720032
Abstract
A linear relationship between particle volume and instrument response is confirmed for particles between 0.75 and 2.5 μ3, but this relationship may not hold for smaller particles. The implicit assumption of negligible surface conductance in derivations of this relationship is found acceptable for polystyrene spheres about 1 μ in diameter. A potential source of error in calibration is identified and allowed for. A relationship is developed between observed count and true count which allows coincident passage correction to be based on all available experimental points. The critical volume is found experimentally to be a function of particle volume.Keywords
This publication has 18 references indexed in Scilit:
- Electrical Counting and Sizing of Mammalian Cells in SuspensionBiophysical Journal, 1965
- Electrical Counting Characteristics of Several Equivolume Micro-organismsJournal of General Microbiology, 1963
- ON THE LOW-FREQUENCY DIELECTRIC DISPERSION OF COLLOIDAL PARTICLES IN ELECTROLYTE SOLUTION1The Journal of Physical Chemistry, 1962
- SIZE DISTRIBUTION OF ERYTHROCYTESAnnals of the New York Academy of Sciences, 1962
- Coincidence in Coulter CountersReview of Scientific Instruments, 1962
- Optical Calibration of Some Monodisperse Polystyrene Latexes in ArraysNature, 1961
- Theory of Coincidence in Coulter Particle CountersReview of Scientific Instruments, 1961
- Changes during the division cycle in bacterial cell wall synthesis, volume, and ability to concentrate free amino acidsBiochimica et Biophysica Acta, 1960
- Electronic Counting and Sizing of BacteriaNature, 1958
- Electron Microscopy of Monodisperse LatexesJournal of Applied Physics, 1955