Inappropriate Ventricular Blanking in a DDI Pacemaker
- 1 July 1986
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 9 (4) , 519-521
- https://doi.org/10.1111/j.1540-8159.1986.tb06608.x
Abstract
The DDI mode is a new pacing mode with potential advantages over DVI pacing. We describe anomalous post R wave ventricular pacing due to the presence of inappropriate ventricular blanking periods in a pacemaker programmed to the DDI mode. Although no adverse consequences were seen in our patients, potentially dangerous R-on-T pacing could occur, particularly if long atrioventricular delays are programmed. A method for eliminating this pacing anomaly is described. Patients programmed to the DDI mode with the pacemaker model described should be evaluated for post R wave ventricular pacing and corrective measures should be taken.Keywords