Osmotic response of lymphocytes measured by means of forward light scattering: Theoretical considerations
Open Access
- 1 November 1988
- Vol. 9 (6) , 636-641
- https://doi.org/10.1002/cyto.990090620
Abstract
Recent data provide solid experimental evidence that increase in cell size of human lymphocytes, caused by osmotic stress, is not directly proportional to forward scattering intensity but rather is inversely proportional. Here, we provide arguments that this phenomenon can be quantitatively explained by means of a model based on a modified Rayleigh‐Debye‐Gans theory. As a consequence, the current view that forward scattering intensities can be used as a measure of gross cell size needs to be reconsidered. In addition, we postulate that structural changes of the cytoplasm, imposed by an osmotic stress, reflect corresponding changes in the nucleus.Keywords
This publication has 19 references indexed in Scilit:
- Light scattering and cell volumes in osmotically stressed and frozen‐thawed cellsCytometry, 1988
- Elastic light scattering from nucleated blood cells: rapid numerical analysisApplied Optics, 1986
- Beam stop and vignetting effects in particle size measurements by laser diffractionApplied Optics, 1986
- Osmotic responses of preimplantation mouse and bovine embryos and their cryobiological implicationsCell Biochemistry and Biophysics, 1986
- Elastic and inelastic light scattering in flow cytometryCytometry, 1983
- Volume regulation by human lymphocytes. Identification of differences between the two major lymphocyte subpopulations.Journal of Clinical Investigation, 1982
- Flow cytometry as an analytical and preparative tool in immunologyJournal of Immunological Methods, 1982
- Dynamics of structural changes in biological particles from rapid light scattering measurementsApplied Optics, 1979
- Light scattering from nucleated biological cellsBiophysical Journal, 1975
- Light Scattering from Coated Spheres: Model for Biological CellsApplied Optics, 1972