Assessment of mean glandular dose in mammography
- 1 August 1995
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 68 (812) , 877-881
- https://doi.org/10.1259/0007-1285-68-812-877
Abstract
The routine assessment of patient dose in the National Health Service Breast Screening Programme is performed as part of the quality assurance protocol recommended by the Institute of Physical Sciences in Medicine. The mean glandular dose to a standard breast is deduced from measurement of the air kerma at the entrance surface of a 4 cm Perspex phantom by applying a series of conversion factors. The exposure factors for this measurement are those used clinically. The measured mean glandular dose is then compared with nationally accepted action levels. In some centres the assessment of mean glandular dose using Perspex is supplemented by patient dose surveys. The mean glandular dose to a series of patients attending a breast screening unit may be estimated from a knowledge of the exposure factors and compressed breast thickness, using a knowledge of the X-ray tube output. Measurements made on units in the Northern Region of England and in Scotland using both methods are presented. The implication of these measurements with regard to patient dose surveys in mammography and quality assurance programmes are discussed. An analysis of the uncertainties associated with the measurement techniques is presented.Keywords
This publication has 4 references indexed in Scilit:
- A method for estimating compressed breast thickness during mammographyThe British Journal of Radiology, 1995
- Perspex blocks for estimation of dose to a standard breast—effect of variation in block thicknessThe British Journal of Radiology, 1995
- Mammographic film density and detection of small breast cancersClinical Radiology, 1994
- Patient dose and risk in mammographyThe British Journal of Radiology, 1991