Normalization of experimental results with respect to inlet conditions in membrane oxygenator testing
- 1 January 1996
- journal article
- research article
- Published by SAGE Publications in Perfusion
- Vol. 11 (1) , 45-51
- https://doi.org/10.1177/026765919601100106
Abstract
This study looked at the problem of the excessive variability in oxygenator testing results, induced by variation of inlet parameters, particularly of inlet oxyhaemoglobin saturation. The investigation was carried out in the laboratory. An in vitro circuit was used to perfuse a small oxygenating cell. Blood flow rate (BFR) and film thickness (BFT) were varied to obtain different oxygenation conditions, while the inspired oxygen fraction (FiO 2) and ventilation ratio were kept at constant values. With each test condition, inlet saturation was varied in the range 60-70% and a number of veno-arterial blood samples (at least 20) were withdrawn and analysed for numerical computing and statistical analysis. The generic law relating oxygenation increment to inlet saturation was found. This allowed a useful normalization procedure to be applied to oxygenator testing results and render them comparable, even if obtained at different inlet conditions.Keywords
This publication has 8 references indexed in Scilit:
- Rethinking the AAMI/ISO 'International Standard' for oxygen transfer performance of artificial lungsPerfusion, 1994
- Standards for blood gas exchange devicesPerfusion, 1994
- Membrane oxygenators: influence of design on performancePerfusion, 1994
- Use of the oxygen transfer slope and estimated membrane oxygen transfer to predict PaO2Perfusion, 1994
- An In Vitro Protocol for Evaluation and Comparison of Membrane OxygenatorsThe Journal of ExtraCorporeal Technology, 1993
- A STANDARDIZED METHOD FOR DEFINING THE OVERALL GAS TRANSFER PERFORMANCE OF ARTIFICIAL LUNGSAsaio Journal, 1972
- Nomograms for correction of blood Po2, Pco2, pH, and base excess for time and temperature.Journal of Applied Physiology, 1966