Plasticizers in P.V.C. and the Occurence of Hefatitis in A Haemodialysis Unit: A Preliminary Communication
- 1 January 1971
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Urology and Nephrology
- Vol. 5 (2) , 141-145
- https://doi.org/10.3109/00365597109133592
Abstract
In a four-year period three isolated cases of hepatitis were observed within a period of 2 months in a haemodialysis unit. The cases occurred in 3 patients during testing of new haemodialysis equipment. Liver biopsy in 2 patients showed non-specific hepatitis and changes as in “viral hepatitis” respectively. From the tested blood tubings, manufactured from polyvinylchloride (P.V.C.), diaethylphthalate was washed out in an amount of 10–20 mg per litre aqueous perfusate. From the tubings normally used in the department this compound was not released on perfusion, and the symptoms disappeared rapidly when treatment with the use of these tubings was resumed.Keywords
This publication has 7 references indexed in Scilit:
- PLASTICISERS FROM P.V.C.The Lancet, 1970
- AUSTRALIA ANTIGEN IN WEST OF SCOTLAND AND NORTH OF ENGLANDThe Lancet, 1970
- CONTAMINATION OF BLOOD STORED IN PLASTIC PACKSThe Lancet, 1970
- An Epidemic of Hepatitis in a Chronic-Hemodialysis UnitNew England Journal of Medicine, 1969
- WHAT'S IN P.V.C.?The Lancet, 1968
- AN ADVERSE EFFECT OF POLYVINYLCHLORIDE TUBING USED IN EXTRACORPOREAL CIRCULATIONThe Lancet, 1968
- Plasticizers in medical application I Analysis and toxicity evaluation of dialkyl benzenedicarboxylatesJournal of Pharmaceutical Sciences, 1967