Cesium Beam Frequency Standard Cavity Design
- 1 April 1970
- journal article
- Published by IOP Publishing in Metrologia
- Vol. 6 (2) , 60-64
- https://doi.org/10.1088/0026-1394/6/2/007
Abstract
A new technique of constructing resonant cavities for exciting the Ramsey resonance in cesium beam frequency standards is discussed. A centrally located E-plane tee is used as the feed system in a symmetric even-order mode cavity. The location of the tee is determined initially from measurements made on the cavity when operated in a symmetric odd-order mode. Both the theoretical conditions and construction methods required to obtain very small phase differences between the oscillating magnetic fields at the ends of the cavity are described. Experimental results indicate agreement with theory and that the method is effective and simple to apply.Keywords
This publication has 1 reference indexed in Scilit:
- A Re-Evaluation of the NRC Long Cesium Beam Frequency StandardMetrologia, 1968