Development of an Implantable Hearing aid Using a Piezoelectric Vibrator of Bimorph Design: State of the Art
- 1 December 1984
- journal article
- research article
- Published by Wiley in Otolaryngology -- Head and Neck Surgery
- Vol. 92 (6) , 706-712
- https://doi.org/10.1177/019459988409200620
Abstract
A new, implantable hearing aid has been developed. An ultraminiature electret microphone placed under the skin of the external ear canal transduces sound waves into electrical impulses that are amplified with battery power. The amplified electrical impulses are fed into a piezoelectric vibrator directly in contact with the stapes. The vibrator transduces the electrical impulses into mechanical vibration with minimal consumption of electrical energy. Direct coupling of the vibrator to the stapes allows a high degree of fidelity in the perception of sound. The developmental process and the structure and function of each component are described together with clinical problems. Preliminary and tentative implantation of this new device during middle ear surgery indicated that it would be beneficial to patients who have suffered hearing loss from middle ear disease and whose condition does not lend itself to surgical correction.Keywords
This publication has 1 reference indexed in Scilit:
- Transformation of sound pressure level from the free field to the eardrum in the horizontal planeThe Journal of the Acoustical Society of America, 1974