Total Heart Replacement with Dual Centrifugal Ventricular Assist Devices
- 1 May 2005
- journal article
- research article
- Published by Wolters Kluwer Health in Asaio Journal
- Vol. 51 (3) , 224-229
- https://doi.org/10.1097/01.mat.0000160400.84250.87
Abstract
In an ovine feasibility study, we implanted two HeartMate-III centrifugal ventricular assist devices (VADs) for total heart replacement. With cardiopulmonary bypass support, both ventricles were transected at the atrioventricular groove, preserving a rim of ventricular tissue. The atrioventricular valves were excised, and the aorta and pulmonary artery were transected above the ventriculoarterial valves. An interatrial septal window was created by excising the foramen ovale. The VADs’ sewing rings were attached to the left and right ventricular remnants, respectively. Outflow grafts were anastomosed to the aorta and pulmonary artery. The left VAD operated continuously at 4,500 rpm. Right VAD speed increased from 2,000 to 4,500 rpm in 500 rpm increments. Outflow graft flow, pressure, oxygen saturation, and shunt direction were recorded. The pulmonary artery to aortic ratio of flow and pressure increased from 0.26 and 0.15 (at 2,000 rpm) to 1.21 and 0.53, respectively (at 4,500 rpm). The interatrial shunt, which was right to left at lower right VAD speeds, progressed to bidirectional, then to left dominant as right VAD speed increased. Outflow-graft oxygen saturation was reflective of the shunt direction. In this acute experiment, total heart replacement with continuous flow VADs satisfactorily balanced left and right ventricular flows and preserved the physiologic circulatory response.Keywords
This publication has 11 references indexed in Scilit:
- HeartMate® VE LVAS design enhancements and its impact on device reliability1European Journal of Cardio-Thoracic Surgery, 2004
- Use of the Flowmaker (Jarvik 2000) Left Ventricular Assist Device for Destination Therapy and Bridging to TransplantationCardiology, 2004
- The MicroMed DeBakey VAD—Part I: The Pump and the Blood FlowThe Journal of ExtraCorporeal Technology, 2003
- Continuous flow pumps and total artificial hearts: management issuesThe Annals of Thoracic Surgery, 2003
- Implantation of biventricular assist devices for chronic heart transplant rejectionThe Annals of Thoracic Surgery, 2002
- HeartMate III: Pump Design for a Centrifugal LVAD with a Magnetically Levitated RotorAsaio Journal, 2001
- Development Status of Terumo Implantable Left Ventricular Assist SystemArtificial Organs, 2001
- Cardiac transplantation: the final therapeutic option for the treatment of heart failureCurrent Opinion in Cardiology, 2000
- IN VITRO AND IN VIVO EVALUATION OF LEFT-RIGHT BALANCING ABILITY OF AN INTERATRIAL SHUNT IN AN ELECTROHYDRAULIC TAH (EHTAH) SYSTEMAsaio Journal, 1997