Quadrature detection surface coil

Abstract
A surface coil assembly consisting of two interleaved coplanar resonators that are intrinsically decoupled from each other is described for imaging at 1.5 T. Vector reception fields on-axis at 3 cm depth are orthogonal and of equal magnitude. Both components of magnetization were received and combined resulting in a 2 1/2 improvement in signal-to-noise for temporomandibular joint images. © 1987 Academic Press, Inc.

This publication has 3 references indexed in Scilit: