Molecular Imaging of Biological Samples: Localization of Peptides and Proteins Using MALDI-TOF MS
- 1 December 1997
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 69 (23) , 4751-4760
- https://doi.org/10.1021/ac970888i
Abstract
Matrix-assisted laser desorption/ionization mass spectrometry (MALDI MS) has been used to generate ion images of samples in one or more mass-to-charge (m/z) values, providing the capability of mapping specific molecules to two-dimensional coordinates of the original sample. The high sensitivity of the technique (low-femtomole to attomole levels for proteins and peptides) allows the study of organized biochemical processes occurring in, for example, mammalian tissue sections. The mass spectrometer is used to determine the molecular weights of the molecules in the surface layers of the tissue. Molecules desorbed from the sample typically are singly protonated, giving an ion at (M + H)+, where M is the molecular mass. The procedure involves coating the tissue section, or a blotted imprint of the section, with a thin layer of energy-absorbing matrix and then analyzing the sample to produce an ordered array of mass spectra, each containing nominal m/z values typically covering a range of over 50 000 Da. Images can be displayed in individual m/z values as a selected ion image, which would localize individual compounds in the tissue, or as summed ion images. MALDI ion images of tissue sections can be obtained directly from tissue slices following preparative steps, and this is demonstrated for the mapping of insulin contained in an islet in a section of rat pancreas, hormone peptides in a small area of a section of rat pituitary, and a small protein bound to the membrane of human mucosa cells. Alternatively, imprints of the tissue can be analyzed by blotting the tissue sections on specially prepared targets containing an adsorbent material, e.g., C-18 coated resin beads. Peptides and small proteins bind to the C-18 and create a positive imprint of the tissue which can then be imaged by the mass spectrometer. This is demonstrated for the MALDI ion image analysis of regions of rat splenic pancreas and for an area of rat pituitary traversing the anterior, intermediate, and posterior regions where localized peptides were mapped. In a single spectrum from the anterior/intermediate lobe of a rat pituitary print, over 50 ions corresponding to the peptides present in this tissue were observed as well as precursors, isoforms, and metabolic frag ments.Keywords
This publication has 16 references indexed in Scilit:
- Matrix-assisted ultraviolet laser desorption of non-volatile compoundsPublished by Elsevier ,2001
- Protein identification with sequence tagsCurrent Biology, 1996
- Sample morphology effects in matrix‐assisted laser desorption/ionization mass spectrometry of proteinsJournal of Mass Spectrometry, 1995
- Improved Resolution and Very High Sensitivity in MALDI TOF of Matrix Surfaces Made by Fast EvaporationAnalytical Chemistry, 1994
- Rapid Communication: Neuropeptide Expression and Processing as Revealed by Direct Matrix‐Assisted Laser Desorption Ionization Mass Spectrometry of Single NeuronsJournal of Neurochemistry, 1994
- Peptide sequencing by matrix‐assisted laser‐desorption mass spectrometryRapid Communications in Mass Spectrometry, 1992
- 2,5-Dihydroxybenzoic acid: a new matrix for laser desorption—ionization mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1991
- Metastable decay of peptides and proteins in matrix‐assisted laser‐desorption mass spectrometryRapid Communications in Mass Spectrometry, 1991
- Removal of Arg1 and Phe22 from CLIP (ACTH18–39) by rodent pituitary and blood peptidasesPeptides, 1988
- Structure of the rat pro‐opiomelanocortin (POMC) geneFEBS Letters, 1985