The hemofiltration/molecular separation (HFMS) artificial kidney concept, first proposed over a decade ago, involves continuous ultrafiltration from the blood stream followed by cleansing of the filtrate, with subsequent return to the body. The system is completely self-contained and portable. During recent preclinical trials on nephrectomized canines, HFMS was better than hemodialysis (HD) in several important ways. The removal or clearance of middle molecules was better with 0.34 m2 HFMS than with 1 m2 HD. Significant phosphate clearance was achieved and the removal rate for creatinine was the same as that for urea. This uniform clearance extends to even higher MW solutes and could potentially result in improved patient response. It mimics the real kidney, whereas membrane-limited dialysis undergoes a logarithmic decrease of clearance with MW. This is due to the fact that solute transport through the membrane involves solution into its matrix followed by diffusion and solute diffusivity decreases with molecular volume. To achieve this potential for hemofiltration-based systems, there are stringent requirements on both the membrane and the plasma proteins allowed to accumulate on the membrane surface.