Shuttle dose at the Vienna Leksell Gamma Knife
- 1 June 1998
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (6) , 1567-1578
- https://doi.org/10.1088/0031-9155/43/6/015
Abstract
The aim of this study was to determine the shuttle dose for all collimator helmets (4, 8, 14 and 18 mm) of the Gamma Knife, model B, in Vienna, Austria. The additional dose accumulated during the transport of the patient in and out of the treatment position should be considered in the dose planning procedure of multicentre treatment regimens and in fractionated stereotactic Gamma Knife radiotherapy. The GafChromic film study was basically used to determine the shuttle dose of all four collimator helmets. In addition, measurements with an ionization chamber (18 and 14 mm collimator) - and, for the 18 mm collimator helmet, TLD dosimetry - were performed in order to confirm the GafChromic film data. The shuttle dose ranged between 99.6 and 183.5 mGy, depending mainly on the size of the collimator and the irradiated isocentres at the half-life activity of Co-60 in a brand new Gamma unit. Our film-generated data were in good correlation with the dose levels obtained with the ionization chamber and the TLD dosimetry, showing a dose difference of less than 0.8%. Since it was possible to verify the shuttle dose even for the 4 and 8 mm collimator helmets, we consider it a non-negligible factor and would advocate the inclusion of the shuttle dose in radiosurgical dose planning.This publication has 13 references indexed in Scilit:
- GafChromic MD‐55: Investigated as a precision dosimeterMedical Physics, 1997
- Dosimetry studies with TLDs for stereotactic radiation techniques for intraocular tumoursPhysics in Medicine & Biology, 1997
- Dosimetric characteristics of an improved radiochromic filmMedical Physics, 1996
- Precision Dosimetry in Narrow Collimated Radiation Beams of the Leksell Gamma KnifeRadiation Protection Dosimetry, 1996
- Transportation Dose and Doses to Extracranial Sites during Stereotactic Radiosurgery with the Leksell Gamma KnifeStereotactic and Functional Neurosurgery, 1996
- The use of radiochromic film in treatment verification of dynamic stereotactic radiosurgeryMedical Physics, 1994
- The use of a radiochromic detector for the determination of stereotactic radiosurgery dose characteristicsa)Medical Physics, 1994
- Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetryMedical Physics, 1991
- Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation and electron beamsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1991
- GafchromicTM dosimetry media: A new high dose, thin film routine dosimeter and dose mapping toolInternational Journal of Radiation Applications and Instrumentation. Part C. Radiation Physics and Chemistry, 1990