A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations*
Top Cited Papers
- 6 April 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (8) , 1767-1790
- https://doi.org/10.1088/0031-9155/50/8/010
Abstract
Based on previous publications on a triple Gaussian analytical pencil beam model and on Monte Carlo calculations using Monte Carlo codes GEANT-Fluka, versions 95, 98, 2002, and BEAMnrc/EGSnrc, a three-dimensional (3D) superposition/convolution algorithm for photon beams (6 MV, 18 MV) is presented. Tissue heterogeneity is taken into account by electron density information of CT images. A clinical beam consists of a superposition of divergent pencil beams. A slab-geometry was used as a phantom model to test computed results by measurements. An essential result is the existence of further dose build-up and build-down effects in the domain of density discontinuities. These effects have increasing magnitude for field sizes < or =5.5 cm(2) and densities < or = 0.25 g cm(-3), in particular with regard to field sizes considered in stereotaxy. They could be confirmed by measurements (mean standard deviation 2%). A practical impact is the dose distribution at transitions from bone to soft tissue, lung or cavities.Keywords
This publication has 32 references indexed in Scilit:
- Analysis of photoneutron spectra produced in medical acceleratorsPhysics in Medicine & Biology, 2000
- Application of an Analytical Pencil Beam Model to Stereotactic Radiation Therapy PlanningJournal of Radiosurgery, 1998
- Radiosurgery for Tumors in the Body: Clinical Experience Using a New MethodJournal of Radiosurgery, 1998
- Calculation of the absorbed dose distribution due to irregularly shaped photon beams using pencil beam kernels derived from basic beam dataPhysics in Medicine & Biology, 1996
- Nuclear Compton scattering in the Δ-resonance region with polarized photonsNuclear Physics A, 1996
- Methods of image reconstruction from projections applied to conformation radiotherapyPhysics in Medicine & Biology, 1990
- Benchmark measurements for lung dose corrections for x-ray beamsInternational Journal of Radiation Oncology*Biology*Physics, 1988
- Optimization of stationary and moving beam radiation therapy techniquesRadiotherapy and Oncology, 1988
- Density functionals for coulomb systemsInternational Journal of Quantum Chemistry, 1983