Comparison of conventional single echo and multi‐echo sequences with a fast spin‐echo sequence for quantitative T2 mapping: Application to the prostate
- 1 July 1996
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 6 (4) , 603-607
- https://doi.org/10.1002/jmri.1880060408
Abstract
The accuracy of water T2 maps generated from a fast spin-echo (FSE) sequence was compared with data obtained by conventional single and multi-echo spin-echo pulse sequences using a commercial gel phantom. Spatially localized stimulated echo acquisition mode (STEAM) proton spectroscopy was also used to confirm the reported water T2 values of the gels contained in the phantom. The FSE sequence was shown to be superior in accuracy to both the single and multi-echo spin echo sequences and comparable to STEAM, producing results that were within 10% of known values. The effectiveness of the FSE sequence was further demonstrated by generating T2 maps of the normal and diseased prostate in clinically acceptable imaging times, resulting in comparable T2 values to those obtained using STEAM. Accurate quantitative T2 maps can be produced with the FSE sequence.Keywords
This publication has 6 references indexed in Scilit:
- Magnetic Resonance Imaging of Prostate CancerTopics in Magnetic Resonance Imaging, 1995
- MR imaging of prostate cancer with an endorectal surface coil technique: correlation with whole-mount specimens.Radiology, 1994
- Quantitation of proton NMR spectra of the human brain using tissue water as an internal concentration referenceNMR in Biomedicine, 1993
- Practical T2 quantitation for clinical applicationsJournal of Magnetic Resonance Imaging, 1992
- Practical choices of fast spin echo pulse sequence parameters: clinically useful proton density and T2-weighted contrastsNeuroradiology, 1992
- Factors influencing contrast in fast spin-echo MR imagingMagnetic Resonance Imaging, 1992