Combined analysis of spatial and velocity displacement artifacts in phase contrast measurements of complex flows
- 1 March 1997
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 7 (2) , 339-346
- https://doi.org/10.1002/jmri.1880070214
Abstract
MR phase contrast (PC) velocity imaging is a promising tool for quantifying blood flow velocity in vivo. PC velocity imaging is, however, susceptible to artifacts that result from the displacement of spins during the finite duration pulse sequences. Such displacement artifacts can lead to errors in velocity measurements, especially in the presence of oblique and accelerating flows, which are common throughout the cardiovascular system. By tracking particles (representing spins) through a computed velocity field, and assuming that spatial and velocity encodings occur at discrete times during the pulse sequence, we simulate the separate and combined effects of oblique and acceleration artifacts on PC velocity images. We demonstrate, both by simulation and MR measurement, the errors associated with such artifacts in PC velocity measurements in a representative flow geometry. Using example particle trajectories, we provide a fluid dynamic basis for characteristic phase‐velocity image distortions that can arise when imaging complex, physiologically relevant flows.Keywords
This publication has 24 references indexed in Scilit:
- Calculation of the magnetization distribution for fluid flow in curved vesselsMagnetic Resonance in Medicine, 1996
- MR measurement and numerical simulation of steady flow in an end-to-side anastomosis modelJournal of Biomechanics, 1996
- Cardiovascular applications of magnetic resonance flow and velocity measurementsJournal of Magnetic Resonance Imaging, 1996
- Determination of wall shear stress in the aorta with the use of MR phase velocity mappingJournal of Magnetic Resonance Imaging, 1995
- Accuracy of MR phase contrast velocity measurements for unsteady flowJournal of Magnetic Resonance Imaging, 1995
- Evaluation of Magnetic Resonance Velocimetry for Steady FlowJournal of Biomechanical Engineering, 1990
- Rapid MR imaging of blood flow with a phase-sensitive, limited-flip-angle, gradient recalled pulse sequence: preliminary experience.Radiology, 1990
- The application of phase shifts in NMR for flow measurementMagnetic Resonance in Medicine, 1990
- Quantitative flow measurements on phantoms and on blood vessels with MRMagnetic Resonance in Medicine, 1988
- The Role of Fluid Mechanics in AtherogenesisJournal of Biomechanical Engineering, 1980