Comparison of adult and paediatric spine and whole body software for the Lunar dual energy X-ray absorptiometer.
- 1 October 1999
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 72 (862) , 967-976
- https://doi.org/10.1259/bjr.72.862.10673948
Abstract
Simple phantoms were devised to compare the performance of adult (software 3.64) and paediatric (software 3.8 g) spine and whole body software developed for the Lunar dual energy X-ray absorptiometer. Rectangular slabs of aluminium with high (1.18 g cm-2) and low (0.57 g cm-2) density were used to represent bone mineral. For spine measurements, the phantoms were scanned in water at depths of 5-20 cm. For whole body measurements, the phantoms were scanned with known amounts of oil and water to represent fat and lean tissue. This simulated tissue depths of 5.5-19.7 cm and body composition ranging from 14-29% fat. There were systematic differences in spine and whole body bone mineral content (BMC), bone area (BA) and bone mineral density (BMD) measurements and also between adult and paediatric software versions. The magnitude and direction of these differences were dependent on BMD of the phantom and tissue depth. Similar systematic differences were observed in vivo when volunteers were scanned using adult and paediatric software. Paediatric software enabled measurements to be made at low tissue depths. The weights of fat, lean and total soft tissue measured by the adult and paediatric whole body software were similar to the values calculated from the known composition of the phantom. Precision estimates for all softwares were excellent. In conclusion, paediatric software should improve bone mineral measurements of children but the discrepancies between adult and paediatric softwares may cause problems in longitudinal studies of skeletal growth and when compiling reference data from infancy through to adulthood.Keywords
This publication has 15 references indexed in Scilit:
- Physical activity and predisposition for hip fractures: A reviewOsteoporosis International, 1997
- Radiation dose and in vitro precision in paediatric bone mineral density measurement using dual X-ray absorptiometry.The British Journal of Radiology, 1997
- Precision and accuracy of measurements of whole-body bone mineral: comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometersThe British Journal of Radiology, 1994
- Consensus development conference: Diagnosis, prophylaxis, and treatment of osteoporosisThe American Journal of Medicine, 1993
- Calcium Supplementation and Increases in Bone Mineral Density in ChildrenNew England Journal of Medicine, 1992
- Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometersThe British Journal of Radiology, 1991
- Role of peak bone mass and bone loss in postmenopausal osteoporosis: 12 year study.BMJ, 1991
- Genetic determinants of bone mass in adult women: A reevaluation of the twin model and the potential importance of gene interaction on heritability estimatesJournal of Bone and Mineral Research, 1991
- Influence of fat on bone measurements with dual-energy absorptiometryBone and Mineral, 1990