Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometers
- 11 November 1991
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 64 (767) , 1023-1029
- https://doi.org/10.1259/0007-1285-64-767-1023
Abstract
The measured absolute bone mineral density values of spine and femur and their precision were compared on two dual energy X-ray absorptiometers, the Lunar DPX and the Hologic QDR-1000. There were systematic differences between the two systems, the Lunar DPX always giving a higher bone density value. The ratio of the mean Lunar DPX/Hologic QDR-1000 bone density measurements obtained in vivo was 1.13 g/cm2 for spine (L2–L4) and 1.20–1.43 g/cm2 for femur measurements. For both systems, short- and long-term precision (coefficient of variation) was below 0.5% for spine in vitro measurements and below 1.5% for neck of femur in vitro measurements. Precision of in vivo measurements was less good, especially for femur measurements. The coefficient of variation of spine measurements was dependent on bone density and deteriorated with decreased bone mineral density and increased body thickness.Keywords
This publication has 13 references indexed in Scilit:
- Interpretation of bone densitometry measurements: Disadvantages of a percentage scale and a discussion of some alternativesJournal of Bone and Mineral Research, 1990
- Dual-energy radiographic absorptiometry of the lumbar spine: clinical experience with two different systems.Radiology, 1990
- Bone Densitometry for Diagnosis and Monitoring OsteoporosisExperimental Biology and Medicine, 1989
- X-ray dual-photon absorptiometry: a new method for the measurement of bone densityThe British Journal of Radiology, 1989
- Performance evaluation of a dual-energy X-ray bone densitometerCalcified Tissue International, 1989
- A theoretical analysis of the accuracy of single-energy CT bone-mineral measurementsPhysics in Medicine & Biology, 1988
- Quantitative Digital RadiographyVersusDual Photon Absorptiometry of the Lumbar Spine*Journal of Clinical Endocrinology & Metabolism, 1988
- Dual energy radiography (DER): A preliminary comparative studyCalcified Tissue International, 1988
- Influence of vertebral fat content on quantitative CT density.Radiology, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986