Comparison of cardiac output measurement techniques: thermodilution, Doppler, GO2‐rebreathing and the direct Fick method
- 1 February 1995
- journal article
- research article
- Published by Wiley in Acta Anaesthesiologica Scandinavica
- Vol. 39 (2) , 245-251
- https://doi.org/10.1111/j.1399-6576.1995.tb04051.x
Abstract
Simultaneously measured cardiac output obtained by thermodilution (TD), transcutaneous suprasternal ultrasonic Doppler (DOP), CO2-rebreathing (CR) and the direct Fick method (FI) were compared in eleven healthy subjects in a supine position (SU), a sitting position (SI), and during sitting exercise at a workload of 50 W (EX). The agreements between the techniques, two by two, were expressed as the bias calculated as the averaged differences between the techniques. Precision was expressed as the standard deviation of the bias. The overall agreement (bias +/- precision) between TD, DOP and CR respectively and FI were 2.3 +/- 1.6, -0.1 +/- 1.4, and -0.2 +/- 1.1 l/min. TD overestimated cardiac output consistently in SU, SI and EX. DOP was in-accurate during EX and agreed well with FI in SU and SI. CR agreed closely with FI in SI and EX, but values were underestimated in SU. The overall agreement between DOP and CR, respectively, and TD were 2.5 +/- 2.2 and 2.6 +/- 1.6 l/min. The overall agreement between DOP and CR was 0.1 +/- 1.6 l/min. In conclusion, TD overestimated cardiac output compared to the other techniques and the poor agreement has to be taken into consideration especially in measures of low values. The precision of DOP and CR against FI seems to be within clinically acceptable limits, and these methods may provide interchangeable alternatives to the invasive Fick method.Keywords
This publication has 26 references indexed in Scilit:
- Automated non-invasive measurement of cardiac output by the carbon dioxide rebreathing method: comparisons with dye dilution and thermodilution.Heart, 1990
- The oxygen status algorithm: A computer program for calculating and displaying pH and blood gas dataScandinavian Journal of Clinical and Laboratory Investigation, 1990
- Acute complications of pulmonary artery catheter insertion in critically ill patientsCritical Care Medicine, 1986
- Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.Heart, 1986
- Analysis of factors affecting the variability of fick versus indicator dilution measurements of cardiac outputThe American Journal of Cardiology, 1985
- HOW FAR IS THE CARDIAC OUTPUT?The Lancet, 1984
- Complications of the flow-directed pulmonary-artery catheterCritical Care Medicine, 1981
- Thermodilution cardiac output: A critical analysis and review of the literatureJournal of Surgical Research, 1979
- A new technique for measurement of cardiac output by thermodilution in manThe American Journal of Cardiology, 1971
- Suprasternal UltrasonographyPublished by American Medical Association (AMA) ,1971