Reproducibility of linear cardiac output measurement by Doppler ultrasound alone.
Open Access
- 1 January 1986
- Vol. 55 (1) , 25-31
- https://doi.org/10.1136/hrt.55.1.25
Abstract
Inclusion of a pig aorta in an artificial circulation with pulsed blood flow allowed correlation of minute distance, measured in the aorta by Doppler ultrasound, and absolute blood flow, measured by timed blood-volume collection. The correlation coefficient was 0.99 with a standard error of prediction that was 5.4% of the minute distance predicted at a standard flow rate of 5 litres per minute. The horizontal distance between 95% confidence limits for a single prediction expressed as a percentage of 5 litres per minute was 33%, and this corresponded to the range of flow rates of 1.65 litres per minute that could give rise to the same measurement. In 142 patients duplicate measurements of minute distance were made with repositioning of the ultrasound transducer between recordings. The mean difference between paired readings, expressed as a percentage of the average (SD) of each pair was 5.4 (4.7)%. Thus, the non-invasive measurement of linear cardiac output by Doppler ultrasound is similarly reproducible in vitro and in vivo and compares favourably with the measurement of volumetric cardiac output by thermodilution.This publication has 22 references indexed in Scilit:
- Assessment of cardiac output by the Doppler ultrasound technique alone.Heart, 1985
- Reproducibility of doppler aortic blood flow measurements: Studies on intraobserver, interobserver and day-to-day variability in normal subjectsThe American Journal of Cardiology, 1984
- Determination of cardiac output by Doppler echocardiography.Heart, 1984
- Doppler echocardiographic measurement of cardiac output: Comparison with a non-golden standardThe American Journal of Cardiology, 1984
- Determination of cardiac output by transcutaneous continuous-wave ultrasonic Doppler computerThe American Journal of Cardiology, 1984
- Noninvasive Doppler determination of cardiac output in man. Clinical validation.Circulation, 1983
- The effect of variations of pulsed Doppler sampling site on calculation of cardiac output: an experimental study in open-chest dogs.Circulation, 1983
- Evaluation of pulmonary and systemic blood flow by 2-dimensional Doppler echocardiography using fast fourier transform spectral analysisThe American Journal of Cardiology, 1982
- Combined doppler and phased-array echocardiographic estimation of cardiac output.Circulation, 1981
- Central Haemodynamics in Essential HypertensionActa Medica Scandinavica, 1977