Combined doppler and phased-array echocardiographic estimation of cardiac output.
- 1 February 1981
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 63 (2) , 388-392
- https://doi.org/10.1161/01.cir.63.2.388
Abstract
The capability of a pulsed Doppler flowmeter combined with a phased-array imaging system to measure volume flow was tested in vitro and in patients undergoing cardiac catheterization. The Doppler-phased-array system (DPA) was used to determine vessel diameter and a superimposed cursor was used to locate the range and angle of the Doppler sample volume. DPA estimates of continuous flow through tubing in a water tank correlated strongly (r = 0.99) with measured flow corresponding to physiologic ranges from 3-12 l/min. For pulsatile flow in a water tank, a correlation of r = 0.86 with measured flow was obtained, whereas DPA estimates of cardiac outputs compared with Fick estimates in the 11 patients produced a correlation of r = 0.83. These data indicate that estimates of cardiac output are possible using the DPA approach.This publication has 8 references indexed in Scilit:
- Synchronized pulsed Doppler blood flow and ultrasonic dimension measurement in conscious dogsUltrasound in Medicine & Biology, 1978
- Color coded ultrasonic differential velocity arterial scanner (Echoflow)Ultrasound in Medicine & Biology, 1978
- An ultrasound system for combined cardiac imaging and Doppler blood flow measurement in man.Circulation, 1978
- Review and Evaluation of Range-Gated, Pulsed, Echo-DopplerJournal of Clinical Engineering, 1978
- Comparison of methods to calculate cardiac output using the CO2 rebreathing methodEuropean Journal of Applied Physiology, 1976
- A New Ultrasound Imaging Technique Employing Two-Dimensional Electronic Beam SteeringPublished by Springer Nature ,1974
- The resolution of the ultrasound pulsed Doppler for blood velocity measurementsJournal of Biomechanics, 1973