Testing the Reliability of a New Ultrasonic Cardiac Output Monitor, the USCOM, by Using Aortic Flowprobes in Anesthetized Dogs
- 1 March 2005
- journal article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 100 (3) , 748-753
- https://doi.org/10.1213/01.ane.0000144774.42408.05
Abstract
We have used an animal model to test the reliability of a new portable continuous-wave Doppler ultrasonic cardiac output monitor, the USCOM. In six anesthetized dogs, cardiac output was measured with a high-precision transit time ultrasonic flowprobe placed on the ascending aorta. The dogs’ cardiac output was increased with a dopamine infusion (0–15 μg · kg−1 · min−1). Simultaneous flowprobe and USCOM cardiac output measurements were made. Up to 64 pairs of readings were collected from each dog. Data were compared by using the Bland and Altman plot method and Lin’s concordance correlation coefficient. A total of 319 sets of paired readings were collected. The mean (±sd) cardiac output was 2.62 ± 1.04 L/min, and readings ranged from 0.79 to 5.73 L/min. The mean bias between the 2 sets of readings was −0.0l L/min, with limits of agreement (95% confidence intervals) of −0.34 to 0.31 L/min. This represents a ±13% error. In five of six dogs, there was a high degree of concordance, or agreement, between the 2 methods, with coefficients >0.9. The USCOM provided reliable measurements of cardiac output over a wide range of values. Clinical trials are needed to validate the device in humans.Keywords
This publication has 16 references indexed in Scilit:
- Impedance CardiographyaChest, 2003
- Electrical bioimpedance for noninvasive measurement of cardiac outputCritical Care Medicine, 1997
- The Effectiveness of Right Heart Catheterization in the Initial Care of Critically III PatientsJAMA, 1996
- Reliability of a New Generation Transesophageal Doppler Device for Cardiac Output MonitoringAnesthesia & Analgesia, 1993
- Thoracic electrical bioimpedance measurement of cardiac output—Not ready for prime timeCritical Care Medicine, 1993
- New perspectives in the assessment of cardiac chamber dimensions during development and adulthoodJournal of the American College of Cardiology, 1992
- Transesophageal Doppler UltrasonographyAnesthesia & Analgesia, 1990
- Noninvasive Cardiac Output: Simultaneous Comparison of Two Different Methods with ThermodilutionAnesthesiology, 1990
- Noninvasive versus invasive assessment of cardiac output after cardiac surgery: clinical validationJournal of Cardiothoracic Anesthesia, 1990
- Noninvasive Doppler determination of cardiac output in man. Clinical validation.Circulation, 1983