Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniques
- 10 October 2000
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 27 (10) , 2198-2208
- https://doi.org/10.1118/1.1289372
Abstract
A two-dimensionally position sensitive dosimetry system has been tested for different dosimetric applications in a radiation therapy facility with a scanning proton beam. The system consists of a scintillating (fluorescent) screen, mounted at the beam-exit side of a phantom and it is observed by a charge coupled device (CCD) camera. The observed light distribution at the screen is equivalent to the two-dimensional (2D)-dose distribution at the screen position. It has been found that the dosimetric properties of the system, measured in a scanning proton beam, are equal to those measured in a proton beam broadened by a scattering system. Measurements of the transversal dose distribution of a single pencil beam are consistent with dose measurements as well as with dose calculations in clinically relevant fields made with multiple pencil beams. Measurements of inhomogeneous dose distributions have shown to be of sufficient accuracy to be suitable for the verification of dose calculation algorithms. The good sensitivity and sub-mm spatial resolution of the system allows for the detection of deviations of a few percent in dose from the expected (intended or calculated) dose distribution. Its dosimetric properties and the immediate availability of the data make this device a useful tool in the quality control of scanning proton beams.Keywords
This publication has 17 references indexed in Scilit:
- An application of GafChromic MD-55 film for 67.5 MeV clinical proton beam dosimetryPhysics in Medicine & Biology, 1999
- A system for three-dimensional dosimetric verification of treatment plans in intensity-modulated radiotherapy with heavy ionsMedical Physics, 1999
- Radiochromic film dosimetry for verification of dose distributions delivered with proton-beam radiosurgeryPhysics in Medicine & Biology, 1997
- Measuring dose distributions for enhanced dynamic wedges using a multichamber detector arrayMedical Physics, 1997
- Strip ionization chambers as 3-D detector for hadron therapyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- The 200‐MeV proton therapy project at the Paul Scherrer Institute: Conceptual design and practical realizationMedical Physics, 1995
- Magnetic scanning system for heavy ion therapyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993
- MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gelsPhysics in Medicine & Biology, 1990
- Film dosimetry of clinical electron beamsInternational Journal of Radiation Oncology*Biology*Physics, 1990
- The response of plastic dosimeters to gamma rays and electrons at high absorbed dose ratesRadiation Physics and Chemistry (1977), 1979