An In Vitro Comparison of Six Microaggregate Blood Filters
- 30 December 1979
- journal article
- research article
- Published by Wiley in Acta Anaesthesiologica Scandinavica
- Vol. 23 (6) , 603-612
- https://doi.org/10.1111/j.1399-6576.1979.tb01493.x
Abstract
The latest models of 6 blood microfilters were tested by filtering large amounts of pooled [human] blood at various driving pressures. For determination of the amount and size of microaggregates in stored blood, a simple method was developed based on consecutive nylon screens. The amount of aggregates was also determined by a modification of the screen filtration pressure method. Filtration efficiency was tested by transfusing blood from the same pool through each of the microfilters. The Pall (SQ40KL) filtered only particles with a diameter greater than 40 .mu.m. Small aggregates also passed through Bentley (PFF-100). The Biotest (MF10B) filter removed large aggregates efficiently, but some small particles were not filtered. Fenwal (4C2423), Swank (6010-3) and Intersept (HRI 8137-00) were the most effective filters. All filters were also tested by pressure transfusion of pooled blood (9 units, 21 days old). The efficiency of Biotest, Pall and especially Bentley improved considerably during the transfusion. Bentley and Fenwal had the highest capacities. A high flow rate was maintained with Pall, but small aggregates were not removed by this filter. Intersept, Swank and Biotest became occluded rapidly. Filtration of 4 units through Intersept released 2218-2454 mg of Hb from the erythrocytes.This publication has 13 references indexed in Scilit:
- In-line Blood Warming and Microfiltration Devices: II. Influence of Blood Temperature on Flow Rate and Hemolysis During Pressure Transfusion Through Microfilters and Transfusion SetsActa Anaesthesiologica Scandinavica, 1979
- Evaluation of the Swank IL-201 Transfusion FilterAnesthesia & Analgesia, 1978
- Counting Microaggregate Particles in BloodTransfusion, 1978
- Effects of Fenwal 4C2423 Transfusion Microfilter on Microaggregates and Other Constituents of Stored BloodTransfusion, 1978
- Acute hemolytic transfusion reactions—a fresh look at pathogenesis and considerations regarding therapyTransfusion, 1977
- Effects of Intersept® Micropore Filtration of Blood on Microaggregates and Other ConstituentsAnesthesiology, 1976
- Pulmonary Microembolism Associated With Massive TransfusionAnnals of Surgery, 1975
- Physical Characteristics of Microaggregates in Stored BloodTransfusion, 1974
- Protection of the Pulmonary Microvasculature by Fine Screen Blood FiltrationChest, 1974
- Prevention of Post-Traumatic Pulmonary InsufficiencyArchives of Surgery, 1973