A Microscale Technique for Gas Chromatography-Mass Spectrometry Measurements of Picogram Amounts of Indole-3-Acetic Acid in Plant Tissues
Open Access
- 1 July 1995
- journal article
- Published by Oxford University Press (OUP) in Plant Physiology
- Vol. 108 (3) , 1043-1047
- https://doi.org/10.1104/pp.108.3.1043
Abstract
A microscale technique has been developed for routine quantifications of picogram amounts of indole-3-acetic acid (IAA) in plant tissues by combined gas chromatography-mass spectrometry. Low- and high-resolution selected-ion-monitoring and selected-reaction-monitoring mass spectrometry techniques were compared for selectivity and precision. The best selectivity was obtained with selected-reaction-monitoring analysis, and 1-mg samples containing 500 fg of IAA could be analyzed accurately with this method. This technique was used to investigate the IAA distribution pattern along the longitudinal axis of tobacco (Nicotiana tabacum [L.]) leaves. In young, developing leaves an increase of endogenous IAA from the leaf tip to the base of the leaf was observed, whereas the level of IAA was uniform along this axis in mature leaves.Keywords
This publication has 7 references indexed in Scilit:
- A Simple Purification of Indole-3-Acetic Acid and Abscisic Acid for GC-SIM-MS Analysis by Microfiltration of Aqueous Samples through NylonPlant Physiology, 1989
- Simultaneous Quantitation of Indole 3-Acetic Acid and Abscisic Acid in Small Samples of Plant Tissue by Gas Chromatography/Mass Spectrometry/Selected Ion MonitoringPlant Physiology, 1987
- Comparison of a Commercial ELISA Assay for Indole-3-Acetic Acid at Several Stages of Purification and Analysis by Gas Chromatography-Selected Ion Monitoring-Mass Spectrometry Using a 13C6-Labeled Internal StandardPlant Physiology, 1987
- 13C6-[Benzene Ring]-Indole-3-Acetic AcidPlant Physiology, 1986
- Dual metastable peak monitoring: Application to the analysis of oestradiol-17β as the bis (tert-butyldimethylsilyl) etherJournal of Mass Spectrometry, 1985
- A comparison of calculation procedures for isotope dilution determinations using gas chromatography mass spectrometryJournal of Mass Spectrometry, 1979
- Selected metastable peak monitoring: A new, specific technique in quantitative gas chromatography mass spectrometryJournal of Mass Spectrometry, 1978