An edge spread technique for measurement of the scatter-to-primary ratio in mammography
- 5 May 2000
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 27 (5) , 845-853
- https://doi.org/10.1118/1.598950
Abstract
An experimental measurement technique that directly measures the magnitude and spatial distribution of scatter in relation to primary radiation is presented in this work. The technique involves the acquisition of magnified edge spread function (ESF) images with and without scattering material present. The ESFs are normalized and subtracted to yield scatter-to-primary ratios (SPRs), along with the spatial distributions of scatter and primary radiation. Mammography is used as the modality to demonstrate the ESF method, which is applicable to all radiographic environments. Sets of three images were acquired with a modified clinical mammography system employing a flat panel detector for 2, 4, 6, and 8 cm thick breast tissue equivalent material phantoms composed of 0%, 43%, and 100% glandular tissue at four different kV settings. Beam stop measurements of scatter were used to validate the ESF methodology. There was good agreement of the mean SPRs between the beam stop and ESF methods. There was good precision in the ESF-determined SPRs with a coefficient of variation on the order of 5%. SPRs ranged from 0.2 to 2.0 and were effectively independent of energy for clinically realistic kVps. The measured SPRs for 2, 4, and 6 cm 0% glandular phantoms imaged at 28 kV were 0.21+/-0.01, 0.39+/-0.01, and 0.57+/-0.02, respectively. The measured SPRs for 2, 4, and 6 cm 43% glandular phantoms imaged at 28 kV were 0.20+/-0.01, 0.35+/-0.02, and 0.53+/-0.02, respectively. The measured SPRs for 2, 4, and 6 cm 100% glandular phantoms imaged at 28 kV were 0.22+/-0.02, 0.42+/-0.03, and 0.88+/-0.08, respectively.Keywords
This publication has 22 references indexed in Scilit:
- Scattered radiation in scanning slot mammographyMedical Physics, 1998
- Performance of glass fiber antiscatter devices at mammographic energiesMedical Physics, 1994
- Scatter in computed radiography.Radiology, 1993
- The effect of scatter reduction on the signal-to-noise ratio in computed radiographyEuropean Journal of Radiology, 1991
- Contrast and scatter in x-ray imaging.RadioGraphics, 1991
- Scatter/primary ratios for x‐ray spectra modified to enhance iodine contrast in screen–film mammographyMedical Physics, 1983
- Grids in mammography.Radiology, 1983
- Analysis of the significance of scattered radiation in reduced dose mammography, including magnification effects, scatter suppression, and focal spot and detector blurringMedical Physics, 1979
- The Intensity of Scattered Radiation in MammographyRadiology, 1978
- X‐ray scatter background signals in transmission radiographyMedical Physics, 1975