Digital selenium radiography: anti-scatter grid for chest radiography in a clinical study.
- 1 September 2000
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 73 (873) , 963-968
- https://doi.org/10.1259/bjr.73.873.11064649
Abstract
The purpose of this study was to evaluate the diagnostic performance of an additional stationary anti-scatter grid in digital selenium radiography (DSR) compared with images acquired with only an air gap. Chest radiographs were obtained with DSR in 100 patients with and without an anti-scatter grid. Four observers scored 12 anatomical landmarks, catheters and wire cerclages for their visualization in both subsets of images. Statistical analysis was performed using a paired t-test. Anatomical landmarks, catheters and wire cerclages were statistically better visualized in regions of high attenuation when the images were performed with an anti-scatter grid. No statistically significant difference was noted for peripheral regions, nor for sex and weight of the patient between the two modalities. Therefore, an anti-scatter grid is not recommended for chest radiography as it increases the radiation exposure of patients without having a significant impact on visualization for all regions of the chest.Keywords
This publication has 8 references indexed in Scilit:
- Detection of Simulated Interstitial Lung Disease and Catheters with Selenium, Storage Phosphor, and Film-based RadiographyRadiology, 1999
- Selenium radiography versus storage phosphor and conventional radiography in the detection of simulated chest lesions.Radiology, 1996
- Chest imaging with a selenium detector versus conventional film radiography: a CT-controlled study.Radiology, 1996
- Phantom Study of Chest Radiography with Storage Phosphor, Selenium, and Film-Screen SystemsActa Radiologica, 1996
- Selenium-based digital radiography of the chest: radiologists' preference compared with film-screen radiographs.American Journal of Roentgenology, 1995
- Technical evaluation of a digital chest radiography system that uses a selenium detector.Radiology, 1995
- Grids or air gaps for scatter reduction in digital radiography: A model calculationMedical Physics, 1992
- Studies of performance of antiscatter grids in digital radiography: Effect on signal‐to‐noise ratioMedical Physics, 1990