Evaluation of a prototype small-animal PET detector with depth-of-interaction encoding
- 16 August 2004
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 51 (4) , 1791-1798
- https://doi.org/10.1109/tns.2004.832905
Abstract
We report on the evaluation of a prototype small-animal positron emission tomography (PET) detector with depth-of-interaction (DOI) encoding. The detector consisted of an 8/spl times/8 array of scintillator crystals that were read out on the top and bottom by position sensitive avalanche photodiodes (PSAPD). Each scintillator crystal had dimensions of 1.65 mm/spl times/1.65 mm/spl times/22.00 mm, and the array had a pitch of 1.75 mm. The 14/spl times/4 mm PSAPDs were coupled to compact, discrete transimpedance preamplifiers. The DOI response was measured by illuminating the detector module from the side with an electronically collimated fan beam of 511 keV gamma rays. The signals from the corner contacts of the PSAPDs were used to identify the crystal of interaction, and the ratio of the total signals from the two PSAPDs was used to calculate the depth z at which the interaction took place. By stepping the crystal array in the z direction (perpendicular to the fan beam), we were able to determine the DOI resolution for each individual crystal. For measurements made at 10/spl deg/C, the average DOI resolution was better than 3 mm. Energy resolution and timing resolution (versus a fast plastic scintillator coupled to a photomultiplier tube) were made under the same operating conditions. The average energy resolution across the array was 16.6% at 511 keV, and the average timing resolution was 3.9 ns. Importantly, the detector performance was maintained all the way out to the crystals at the edges of the PSAPDs.Keywords
This publication has 10 references indexed in Scilit:
- Position Sensitive APDs for Small Animal PET ImagingIEEE Transactions on Nuclear Science, 2004
- Evaluation of a position sensitive avalanche photodiode for PETIEEE Transactions on Nuclear Science, 2003
- Position-sensitive avalanche photodiodes for gamma-ray imagingIEEE Transactions on Nuclear Science, 2002
- Dual APD array readout of LSO crystals: optimization of crystal surface treatmentIEEE Transactions on Nuclear Science, 2002
- Properties of avalanche photodiodes for applications in high energy physics, astrophysics and medical imagingNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- Performance characteristics of micro crystal element (MiCE) detectorsIEEE Transactions on Nuclear Science, 2001
- An LSO scintillator array for a PET detector module with depth of interaction measurementIEEE Transactions on Nuclear Science, 2001
- Geometry and surface treatment dependence of the light collection from LSO crystalsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- Gain and noise in very high-gain avalanche photodiodes: theory and experimentPublished by SPIE-Intl Soc Optical Eng ,1996
- Design studies for a PET detector module using a PIN photodiode to measure depth of interactionIEEE Transactions on Nuclear Science, 1994