Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements
Top Cited Papers
- 1 January 2004
- journal article
- research article
- Published by Springer Nature in European Journal of Applied Physiology
- Vol. 91 (1) , 22-29
- https://doi.org/10.1007/s00421-003-0933-0
Abstract
This study quantitatively assessed the mechanical reliability and validity of position, torque and velocity measurements of the Biodex System 3 isokinetic dynamometer. Trial-to-trial and day-to-day reliability were assessed during three trials on two separate days. To assess instrument validity, measurement of each variable using the Biodex System 3 dynamometer was compared to a criterion measure of position, torque and velocity. Position was assessed at 5° increments across the available range of motion of the dynamometer. Torque measures were assessed isometrically by hanging six different calibrated weights from the lever arm. Velocity was assessed (30°/s to 500°/s) across a 70° arc of motion by manually accelerating the weighted lever arm. With the exception of a systematic decrease in velocity at speeds of 300°/s and higher, the Biodex System 3 performed with acceptable mechanical reliability and validity on all variables tested.Keywords
This publication has 9 references indexed in Scilit:
- Inertial Effects on Moment Development During Isokinetic Concentric Knee Extension TestingJournal of Orthopaedic & Sports Physical Therapy, 2000
- Prerequisites and limitations to isokinetic measurements in humansEuropean Journal of Applied Physiology, 1996
- The Utility of Isokinetic Dynamometry in the Assessment of Human Muscle FunctionSports Medicine, 1996
- Validity and Reliability of the LIDO Active Isokinetic SystemJournal of Orthopaedic & Sports Physical Therapy, 1992
- Static and dynamic assessment of the Biodex dynamometerEuropean Journal of Applied Physiology, 1991
- Test-Retest Reliability of the Biodex Isokinetic DynamometerJournal of Orthopaedic & Sports Physical Therapy, 1990
- CONSTANT VELOCITY DYNAMOMETER - AN APPRAISAL USING MECHANICAL LOADING1986
- Analysis of the reliability and validity of the kinetic communicator exercise deviceMedicine & Science in Sports & Exercise, 1986
- The nature of torque ???overshoot??? in Cybex isokinetic dynamometryMedicine & Science in Sports & Exercise, 1982