Repeatability of corneal thickness measurements made by a scanning slit topography system
- 1 November 2002
- journal article
- research article
- Published by Wiley in Ophthalmic and Physiological Optics
- Vol. 22 (6) , 505-510
- https://doi.org/10.1046/j.1475-1313.2002.00060.x
Abstract
Two sets of corneal thickness (CT) measurements at five corneal locations were made with Orbscan on 33 normal young adults. No significant intravisit and intervisit differences at any of the corneal locations were found (Repeated measures anova, F < 0.49, p > 0.49). The 95% limits of agreement (95% LA) for between‐measurement differences for the central cornea was about ±19 μm (or 3%), indicating good repeatability; for the inferior, temporal and nasal corneal locations, the 95% LA were close to ±30 μm (≈4–5%); and for the superior corneal location, they were about ±35 μm (or 5.5%). The estimated number of repeated measurements that should be taken for accurate CT measurements (at any of the five corneal locations and excluding outliers of >3 S.D.) of 2% (standard error) is ≤2, and of 3% is 1. This study shows that only central CT measurements with the Orbscan are repeatable. The role of the Orbscan pachometry is therefore limited and is recommended for central CT measurements only.Keywords
This publication has 21 references indexed in Scilit:
- Factors affecting the forward shift of posterior corneal surface after laser in situ keratomileusisOphthalmology, 2001
- The Repeatability of Corneal Thickness MeasuresCornea, 2000
- Correlation between intraocular pressure and corneal stromal thickness after laser in situ keratomileusisJournal of Cataract & Refractive Surgery, 2000
- Orbscan pachymetry: Implications of a repeated measures and diurnal variation analysisOphthalmology, 1999
- Discussion by Robert K. Maloney, MDOphthalmology, 1999
- Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness11The authors have no propriety interests in any materials or methods described in this article.Ophthalmology, 1999
- Changes in refractive trends and optical components of Hong Kong Chinese aged 19–39 yearsOphthalmic and Physiological Optics, 1994
- Relative Sensitivity of Clinical Tests to Hydrophilic Lens- Induced Corneal Thickness ChangesOptometry and Vision Science, 1993
- Topographical corneal edemaActa Ophthalmologica, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986