The Repeatability of Corneal Thickness Measures
- 1 November 2000
- journal article
- research article
- Published by Wolters Kluwer Health in Cornea
- Vol. 19 (6) , 792-795
- https://doi.org/10.1097/00003226-200011000-00007
Abstract
To compare the repeatability of three measures of corneal thickness: Orbscan Slitscan pachymetry, ultrasound pachymetry, and optical pachymetry. Twenty normal subjects were tested on three occasions. Two occurred on the same day and the third was on a different day at approximately the same time of day as one of the first two visits. Central corneal thickness of the right eye was measured with a Haag-Streit optical pachymeter, a Humphrey Model 855 ultrasound pachymeter, and the Orbscan system. Day-to-day and same-day repeatability was assessed by calculating the difference between the values from two visits and determining the mean difference, the SD, and the 95% limits of agreement (LoA) (LoA = mean +/- 1.96 SD). Mean (+/- SD) central corneal thickness as measured by each instrument was as follows: 539 +/- 33 microm for optical pachymetry, 542 +/- 33 microm for ultrasound pachymetry, and 596 +/- 40 microm for Orbscan pachymetry. For day-to-day comparisons, optical pachymetry showed the poorest repeatability with 95% LoA of -61 to +32 microm. Ultrasound pachymetry showed better repeatability with 95% LoA of -22 to +24 microm. The Orbscan showed the best repeatability centrally with 95% LoA of -10 to +17 microm. Peripheral Orbscan pachymetry was less repeatable than that measured centrally but still more repeatable than central optical pachymetry. Similar results were found with same-day comparisons. The Orbscan system is the most repeatable technique for measuring corneal thickness but shows a significant bias toward greater corneal thickness measures than both ultrasound and optical pachymetry.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography systemBritish Journal of Ophthalmology, 1999
- Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness11The authors have no propriety interests in any materials or methods described in this article.Ophthalmology, 1999
- Corneal Response to OrthokeratologyOptometry and Vision Science, 1998
- Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetryJournal of Cataract & Refractive Surgery, 1997
- Extended Wear in PerspectiveOptometry and Vision Science, 1997
- The effect of age on human corneal thicknessActa Ophthalmologica, 1993
- Variability of Ultrasonic PachometryOptometry and Vision Science, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Optical Pachometry MeasurementsOptometry and Vision Science, 1984
- PachometryOptometry and Vision Science, 1983