Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis
Open Access
- 1 April 2004
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 88 (4) , 466-468
- https://doi.org/10.1136/bjo.2003.030361
Abstract
Aims: To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). Methods: Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. Results: Three devices gave significantly different corneal thickness readings (pr = 0.912, pr = 0.968, pr = 0.933, pConclusion: In post-LASIK eyes, Orbscan II scanning slit topography significantly underestimated corneal thickness. Non-contact specular microscopy gave smaller thickness readings than ultrasonic pachymetry, but these two units showed an excellent linear correlation.Keywords
This publication has 17 references indexed in Scilit:
- Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetryEye, 2002
- A comparison of two pachymetric systems: slit-scanning and ultrasonic.Published by Rockefeller University Press ,2002
- Comparison of corneal pachymetry using ultrasound and Orbscan IIJournal of Cataract & Refractive Surgery, 2002
- Accuracy of Orbscan optical pachymetry in corneas with hazeJournal of Cataract & Refractive Surgery, 2002
- Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyesJournal of Cataract & Refractive Surgery, 2001
- Scanning-slit and Specular Microscopic Pachymetry in Comparison With Ultrasonic Determination of Corneal ThicknessCornea, 2001
- Corneal thickness measurements with contact and noncontact specular microscopic and ultrasonic pachymetryAmerican Journal of Ophthalmology, 2001
- Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement proceduresJournal of Cataract & Refractive Surgery, 2001
- Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography systemBritish Journal of Ophthalmology, 1999
- Corneal Thickness Measurements With the Topcon SP-2000P Specular Microscope and an Ultrasound PachymeterArchives of Ophthalmology (1950), 1999