Corroded Nitinol Wires in Explanted Aortic Endografts: An Important Mechanism of Failure?
- 1 June 2001
- journal article
- research article
- Published by SAGE Publications in Journal of Endovascular Therapy
- Vol. 8 (3) , 248-253
- https://doi.org/10.1177/152660280100800303
Abstract
Purpose: To analyze surface alterations and fractures observed in the nitinol stent wires of explanted endovascular grafts used for treatment of abdominal aortic aneurysm. Methods: Twenty-one explanted Stentor devices and 1 Cragg stent were received from investigators in Germany and France. After macroscopy and photography, the explants were cleaned and the polyester coating removed. The frame was examined completely by stereomicroscopy, and irregularities were assayed by scanning electron microscopy and energy dispersive x-ray analysis (EDAX). The observed alterations were classified according to stereomicroscopic and electron microscopic morphology. Results: The mean implantation interval for the endografts was 29.1 ± 13.2 months (range 5–46). All examined explants, even those retrieved after only a few months in situ, showed pitlike surface damage 10–25 µm in diameter. Larger, irregularly shaped surface alterations were observed in ∼70% of the explants. Older explants (age >32 months) presented vast regions of decay, with bending of the wire and stress cracks in some areas. EDAX examination revealed decreased nickel concentration in the corroded regions. Conclusions: Corrosion of the nitinol wire in endovascular grafts is confirmed. Presumably, the observed pitting and irregularly shaped corrosion defects are the precursors of material failure. They weaken the thin wire, which leads to stress cracks and eventually fracture of the stent wire under circulatory pulsation. Cell-induced electrochemical corrosion and active cellular destruction of surfaces are well-known mechanisms that must be investigated for their possible roles in the corrosion of stent metals.Keywords
This publication has 8 references indexed in Scilit:
- First-Generation Aortic Endografts: Analysis of Explanted Stentor Devices from the EUROSTAR RegistryJournal of Endovascular Therapy, 2000
- Frame Dislocation of Body Middle Rings in Endovascular Stent Tube GraftsEuropean Journal of Vascular and Endovascular Surgery, 1999
- Biocompatibility of nickel-titanium shape memory metal and its corrosion behavior in human cell culturesJournal of Biomedical Materials Research, 1997
- Improvement of the Corrosion Resistance of NiTi Stents by Surface TreatmentsMRS Proceedings, 1996
- Microbially Influenced Corrosion of MaterialsPublished by Springer Nature ,1996
- The Pathology of Vascular GraftsPublished by Springer Nature ,1994
- The corrosion behaviour of nickel titanium shape memory alloysCorrosion Science, 1990
- Biocompatibility of nitinol alloy as an implant materialJournal of Biomedical Materials Research, 1976