Correction factors for water‐proofing sleeves in kilovoltage x‐ray beams
- 1 September 1997
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 24 (9) , 1507-1513
- https://doi.org/10.1118/1.598039
Abstract
This paper investigates the effect of the waterproofing sleeve on the calibration of kilovoltage photon beams (50–300 kV). The sleeve effect correction factor,has been calculated using the Monte Carlo method as the ratios of the air kerma in an air cavity of a cylindrical chamber without the waterproofing sleeve to that with a sleeve. Three sleeve materials have been studied, PMMA, nylon and polystyrene. The calculations were carried out using the EGS4 (Electron Gamma Shower version 4) code system with the application of a correlated‐sampling variance‐reduction technique. The results show that the sleeve correction factor for a 1‐mm thick PMMA waterproofing sleeve surrounding a Farmer‐type cylindrical chamber at 2 cm depth in water is about 0.991 at 50 kV (1.04 mm Al HVL) and 1.000 at 300 kV (21.5 mm Al HVL). For 1‐mm thick nylon and polystyrene sleeves,varies from 0.992 to 1.000 and from 0.981 to 1.000, respectively, for the same beam quality range. Thefactor varies with sleeve thickness, beam quality and phantom depth. No significant dependence of thefactor on field size and source‐surface distance has been found. Measurements for PMMA, nylon and polystyrene sleeves of various thicknesses have also been carried out and show excellent agreement with the Monte Carlo calculations.Keywords
This publication has 13 references indexed in Scilit:
- The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al - 4 mm Cu HVL; 10 - 300 kV generating potential)Physics in Medicine & Biology, 1996
- Dependence of overall correction factor of a cylindrical ionization chamber on field size and depth in medium‐energy x‐ray beamsMedical Physics, 1996
- Monte Carlo calculated stem effect corrections for NE2561 and NE2571 chambers in medium-energy X-ray beamsPhysics in Medicine & Biology, 1995
- Calculations of ion chamber displacement effect corrections for medium-energy X-ray dosimetryPhysics in Medicine & Biology, 1995
- Calculation of absorbed dose ratios using correlated Monte Carlo samplingMedical Physics, 1993
- Correction factors for a cylindrical ionization chamber used in medium-energy X-ray beamsPhysics in Medicine & Biology, 1993
- The effect of waterproofing sleeves on ionization chamber responsePhysics in Medicine & Biology, 1992
- Determination of absorbed dose to water with ionisation chambers calibrated in free air for medium-energy X-raysPhysics in Medicine & Biology, 1988
- Effects of plastic protective caps on the calibration of therapy beams in waterMedical Physics, 1985
- The effect of thickness of the waterproofing sheath on the calibration of photon and electron beamsMedical Physics, 1985