Acute and Long‐Term Atrial and Ventricular Stimulation Thresholds with a Steroid‐Eluting Electrode
- 1 January 1985
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 8 (1) , 45-49
- https://doi.org/10.1111/j.1540-8159.1985.tb05721.x
Abstract
Thirty leads with a steroid‐eluting electrode (Medtronic 4003 and 4503) have been implanted in 24 patients, 11 in the atrium and 19 in the ventricle. Six patients received the steroid lead in both atrium and ventricle. The stimulation thresholds were followed using Elema pulse generators with Vario function in 15 patients (11 atrial leads and 10 ventricular leads) during 11 ± 3 months (mean ± SD). At 0.5 ms pulse duration the mean atrial and ventricular thresholds were 0.55 Volt ± 0.22 and 0.39 Volt ± 0.22 respectively at implant, 0.94 Volt ± 0.13 and 0.82 Volt ± 0.16 after 12 months. During the entire follow‐up period both atrial and ventricular mean thresholds never exceeded 1 V.Keywords
This publication has 5 references indexed in Scilit:
- A New Steroid-Eluting Low Threshold Pacemaker LeadPublished by Springer Nature ,1983
- Modern atrial and ventricular leads for permanent cardiac pacing.Heart, 1981
- A New Lead for Transvenous Atrial Pacing and Sensing Clinical and Electrophysiological ExperiencesPacing and Clinical Electrophysiology, 1980
- Influence of Output Capacitor, Electrode and Pulse Width on Power Consumption in Cardiac Pacing‡Pacing and Clinical Electrophysiology, 1978
- ALTERATION OF PACEMAKER THRESHOLD BY DRUG AND PHYSIOLOGICAL FACTORSAnnals of the New York Academy of Sciences, 1969