Polymeric Surface Modifications of Tantalum Stents
- 1 August 1996
- journal article
- research article
- Published by SAGE Publications in Journal of Endovascular Therapy
- Vol. 3 (3) , 276-283
- https://doi.org/10.1177/152660289600300306
Abstract
Purpose: To compare two kinds of polymer-coated tantalum stents with bare tantalum stents (control) to determine if the coatings can improve thromboresistance. Methods: Twenty-seven Fontaine-Dake stents were balloon expanded in three 8-mm × 80-cm.polytetrafluoroethylene (PTFE) grafts; 9 stents were bare tantalum (T); 9 were coated with polyetherurethane (PL); and 9 were coated with parylene (PA). There were 9 stents placed in each graft as follows: 3 tantalum, 3 polyetherurethane, and 3 parylene. In swine whose platelets had been radiolabeled with indium 111, the ends of each stented graft were connected to 14F femoral and venous sheaths to create an ex vivo fistula. Each graft was exposed to blood for 30, 60, and 120 minutes. At the end of each test period, the stented grafts were disconnected from the sheaths, flushed with saline until clear, and then flushed with formalin. The stents were removed from the grafts, and a radionuclide well counter recorded radionuclide counts from each stent type at each period of blood contact. These values were converted to platelet density per 1000 mUm2. Stents were then photographed and scanned with electron microscopy (EM) for qualitative analysis. Possible significant differences in platelet adhesion with the three types of stents (both between stent groups and within stent groups) were examined using a two-tailed Student's f-test. Results: There were significantly fewer platelets adsorbed on PA versus T at all time periods (p < 0.005); on PL versus T at 60 and 120 minutes (p < 0.005); and on PA versus PL at 30 and 120 minutes (p < 0.0005). There was no significant difference in platelet density within each stent group (p = 0.1). Mean platelet density (number of platelets per 1000 mUm2 ± SD) was as follows: at 30 minutes: T = 1891 ± 965; PL = 373 ± 193; and PA = 27 ± 3; at 60 minutes: T = 6226 ± 1621; PL = 1573 ± 793; and PA = 1185 ± 710; at 120 minutes: T = 5307 ± 591; PL = 3164 ± 318; and PA = 180 ± 100. Gross inspection of the 120-minute groups demonstrated focal areas of thrombus on T, less on PL, and none on PA. Scanning EM demonstrated extensive platelet accumulation covering T at all time periods, less on PL, and even less on PA. Conclusions: Polymeric surface modification of tantalum stents with parylene and/or polyetherurethane can improve the acute thromboresistance of these devices; parylene appears to be the more thromboresistant of the two coatings.Keywords
This publication has 10 references indexed in Scilit:
- Use of Polyetherurethane to Improve the Biocompatibility of Vascular StentsJournal of Endovascular Therapy, 1995
- Decreased Platelet Adherence of Polymer-coated Tantalum StentsJournal of Vascular and Interventional Radiology, 1994
- Intravascular stents: tissue-stent interactions and design considerations.American Journal of Roentgenology, 1993
- Blood compatibility of PEO grafted polyurethane and HEMA/styrene block copolymer surfacesJournal of Biomedical Materials Research, 1990
- Long-term implants of Parylene-C coated microelectrodesMedical & Biological Engineering & Computing, 1988
- Preparation and properties of plasma-deposited films with surface energies varying over a wide rangeJournal of Applied Polymer Science, 1986
- Polyether—urethane ionomers: surface property/ex vivo blood compatibility relationshipsJournal of Colloid and Interface Science, 1985
- Bioevaluation of plasma polymerized films in skeletal muscleJournal of Biomedical Materials Research, 1979
- Growth of cultured calf aortic smooth muscle cells on cardiovascular prosthetic materialsJournal of Biomedical Materials Research, 1976
- Ultrathin Microfiber Lining for Artificial OrgansTextile Research Journal, 1973